Saga Cryogenic

Saga Cryogenic: Cryogenic monitoring down to −200 °C

Built for cell and gene therapies shipped in liquid nitrogen: the probe reads down to −200 °C inside the shipper, while the Saga Cryogenic device rides outside it.

Saga Cryogenic device, white at the top and orange from the screen down, PR-T80 probe attached and coiled below it, screen reading −195.7 °C
Saga Cryogenic, 25 × 61 × 98 mm, 194 g. Drawn to scale on the housing's 61 mm width beside a credit card (ISO/IEC 7810 ID-1).Source: Saga Cryogenic datasheet, with the PR-T80 probe
Operating range
−200 to +50 °C
Through the PR-T80 probe
Accuracy
±1.4 °C
at −200 to −90 °C
Housing
61 × 98 mm
25 mm deep, 194 g
IP rating
IP54

Key capabilities

  1. −200 °C operation

    The PR-T80 probe, calibrated at −196, −80, −40 and 0 °C, monitors liquid-nitrogen shipments in real time.

  2. Cell and gene therapy ready

    Location, temperature, and light tracking for the most valuable therapies.

  3. Validated for the life sciences supply chain

    Validated monitoring with a complete audit trail, on the same platform as the rest of the lane.

Specifications

The same rows as every Saga, read from the datasheet

version 2 · 7 April 2026 · 331 KB. Where a figure is missing from the sheet the row says so rather than borrowing from another model.

Limited range

The probe reads to −200 °C. The device itself rides outside the liquid nitrogen at −20 to +50 °C; between −30 and −20 °C it still logs, with reduced battery life, but stops reporting in real time.

Specifications
ModelSaga Cryogenic
Standard operating range−200 to +50 °C
ProbePR-T80 digital temperature probe
Calibration points−196, −80, −40, 0 °C on the probe, NIST traceable
Accuracy±1.4 °C at −200 to −90 °C · ±1 °C at −90 to 0 °C
Resolution · long-term drift0.1 °C · under 0.1 °C per year
Logging · reportingLogs every 10 minutes · reports in real time every 1 to 24 hours, and on any excursion
Battery2500 mAh rechargeable NiMH, not dangerous goods, at least 70 charge cycles
Memory150 days of readings at 10-minute intervals
Connectivity · location4G, 3G and 2G, global · cellular triangulation and 2.4 GHz Wi-Fi detection
Dimensions · weight25 × 61 × 98 mm · 194 g
Housing · ingress protectionABS · IP54
ApprovalsCE, FCC, IC, NOM, KC · IATA, FAA and EASA flight compliance

Battery life

Set the wake-up interval, read the days

A fully charged device on the interval shown reports in real time and raises alarms for at least this long while monitoring a shipment inside its operating range. Below 5% charge it hibernates: it keeps logging to memory and stops reporting until recharged.

Battery life by wake-up interval
Wake-up interval2500 mAh
1 hour20 days
2 hours35 days
3 hours45 days
6 hours75 days
12 hours110 days
24 hoursover 110 days

In the cold

A deep-freeze chamber controller reading minus 51.320°C against a set point of minus 81°C, with probe harnesses behind it
Deep-freeze validation in Reykjavik: −51.320°C on the way to a −81°C set point.

For the most extreme cold chains

Cell and gene therapies travel in liquid nitrogen, and until Saga Cryogenic, monitoring them in real time was nearly impossible. The reusable device tracks temperature, location, and light exposure with full GxP compliance, and the cryogenic probe has an operating range from −200 °C to +50 °C.

That range covers cryogenic transport and cryogenic storage, so traceability, regulatory compliance, and product safety hold across both rather than stopping at the loading dock.

Device management is handled

Return logistics, calibration, and servicing are ours to run, which keeps a device and its probe in circulation rather than in a drawer.

Questions about Saga Cryogenic

How cold can Saga Cryogenic measure?

The PR-T80 probe reads to −200 °C. It is calibrated at −196, −80, −40 and 0 °C on the probe, and accurate to ±1.4 °C from −200 to −90 °C and ±1 °C from −90 to 0 °C.

Where does the device sit in a liquid nitrogen shipper?

Outside it, in its own −20 to +50 °C band, with only the probe in the cold. Between −30 and −20 °C the device still logs, with reduced battery life, but stops reporting in real time.

Is it the same device as Saga Dry Ice?

Same Saga-P housing and the same PR-T80 probe. What differs is the calibration and the accuracy bands: Cryogenic is calibrated at −196, −80, −40 and 0 °C on the probe and reads to −200 °C; Dry Ice stops at a −80 °C point and reads to −95 °C.