Saga-P Standard
Saga-P Standard: +2 to +8 °C and room temperature, one device
Saga-P Standard covers −20 to +50 °C, calibrated at −20, +15, and +50 °C and accurate to ±0.5 °C across the refrigerated and ambient bands.

- Operating range
- −20 to +50 °C
- Accuracy
- ±0.5 °C
- at −10 to +45 °C
- Housing
- 61 × 103 mm
- 25 mm deep, 197.5 g
- IP rating
- IP55
Key capabilities
Refrigerated and ambient in one range
±0.5 °C from −10 to +45 °C, which covers +2 to +8 °C and controlled room temperature without a second device.
110 days on a 12-hour wake-up
A 2500 mAh NiMH cell, not dangerous goods, recharged over the attached USB-A cable for at least 70 cycles.
Alerts the moment it leaves range
Reports an excursion as it happens, regardless of the wake-up schedule, so you can act while the shipment is still on its way.
Specifications
Built for +2 to +8 °C and room temperature
Full specifications from the Saga-P Standard datasheet, version 1.0, 18 March 2026.
Limited range
Between −30 and −20 °C, and from +50 to +70 °C, the device still logs. Battery life is reduced, the display is not functional, and below −20 °C it stops reporting in real time.
| Model | Saga-P Standard |
|---|---|
| Standard operating range | −20 to +50 °C |
| Probe | None |
| Calibration points | −20, +15, +50 °C, NIST traceable |
| Accuracy | ±0.5 °C at −10 to +45 °C · ±1 °C outside that range |
| Resolution · long-term drift | 0.1 °C · under 0.1 °C per year |
| Logging · reporting | Logs every 10 minutes · reports in real time every 1 to 24 hours, and on any excursion |
| Battery | 2500 mAh rechargeable NiMH, not dangerous goods, at least 70 charge cycles |
| Memory | 150 days of readings at 10-minute intervals |
| Connectivity · location | 4G, 3G, and 2G, global · cellular triangulation and 2.4 GHz Wi-Fi detection |
| Dimensions · weight | 25 × 61 × 103 mm · 197.5 g |
| Housing · ingress protection | ABS · IP55 |
| Approvals | CE, FCC, IC, NOM, KC · IATA, FAA, and EASA flight compliance |
Battery life
Set the wake-up interval, read the days
A fully charged device on the interval shown reports in real time and raises alarms for at least this long while monitoring a shipment inside its operating range. Below 5% charge it hibernates: it keeps logging to memory and stops reporting until recharged.
| Wake-up interval | 2500 mAh |
|---|---|
| 1 hour | 20 days |
| 2 hours | 35 days |
| 3 hours | 45 days |
| 6 hours | 75 days |
| 12 hours | 110 days |
| 24 hours | over 110 days |
Device management is handled
Return logistics, calibration, and servicing are ours to run. You get maximum reuse and a circular model without adding a device operation to your own team.
Questions about Saga-P Standard
What does the P in Saga-P mean?
The PR-T80 probe port. It ships capped, so on a refrigerated or ambient lane the Saga-P Standard reads temperature at the housing, with no probe to handle. The datasheet states 25 × 61 × 103 mm and 197.5 g.
Does one device cover both +2 to +8 °C and controlled room temperature?
Yes. Saga-P Standard covers −20 to +50 °C and is accurate to ±0.5 °C from −10 to +45 °C, which spans the refrigerated and ambient bands without a second device. It is calibrated at −20, +15, and +50 °C.
What happens if the shipment goes below −20 °C?
Between −30 and −20 °C, and from +50 to +70 °C, the device still logs. Battery life is reduced, the display is not functional, and below −20 °C it stops reporting in real time. For product that ships frozen as a matter of course, Saga Arctic is calibrated for it.
Put two Saga devices on your own shipments
Controlant Go sends two pre-configured Saga devices, free of charge, with nothing to implement.